Ny innovativ teknik att frysa in vävnad, inga farliga kemikalier används.
The complete system includes the following:
Cryoembedding unit for embedding of fresh tissues for frozen sections procedures.
Face down embedding technique, US Patent 6,536,219 B2
Standardized freezing at -40°C for all types of tissues
Rapid change out molds with a range of different sizes
Only 40 minutes from room temperature to operating temperature of -40°C
N. 4 moulds with independent timer
Optimum planarity of specimens mounted in cryoembedding compound for reduced trimming time
Advanced technology Stirling cooler with Helium gas as a refrigerant in sealed stainless steel chamber. No CFC (Chlorofluorocarbon)
USB port for downloading of event logs and software updating
Fully automatic defrost program
Touch screen terminal 4,3"
Unit includes:
Mould diameter 30mm (h 5mm) (Q.ty 2)
Mould diameter 40mm (h 5mm) (Q.ty 2)
Heat Extractor (Q.ty 2)
Small tip spatula (Q.ty 1)
Large tip spatula (Q.ty 1)
Tweezer Inox (110mm) (Q.ty 1)
Tweezer Inox (140mm) (Q.ty 1)
Box of lens paper disks, diameter 22mm (100 pieces) (Q.ty 1)
Tool for mold exchange (Q.ty 1)
Kit Hepa filter
Dimensions: L 300mm/11,8", D 540mm/21,3", H 450mm/17,7" (730mm/28,7" with cover open)
Weight: 25 kg/56lb
Power supply: 230 Volt 50/60Hz 250W
OBS! Byte av HEPA filter. Informera kunden om detta:
Every 250 defrost pump hours
For safety reasons, between the working area and the defrost pump, there is a HEPA12 filter that filters
possibly dangerous pathogens in the vapors.
This filter must be changed after the defrost pump has worked for 250 hours.
The software advises you that the filter life is finished with the message “Filter needs to be replaced. Please
do it”, at the starting of the first freezing program of the day.'
//linda 180501